Skip to main content

Tandem’s Control-IQ System Increases Time in Range and Lowers A1C in People with Type 1 Diabetes

Updated: 8/14/21 1:00 amPublished: 6/28/19
By Jeemin KwonAdam Brown

By Jeemin Kwon and Adam Brown

The automated insulin delivery system with Dexcom’s G6 aims to launch in the US by the end of the year for those ages 14+; first system with auto boluses and no fingerstick calibration

At the American Diabetes Association’s (ADA) 79th Scientific Sessions, researchers shared highly anticipated results from Tandem’s Control-IQ hybrid closed loop pivotal trial. Control-IQ is an automated insulin delivery (AID) system comprised of the t:slim X2 insulin pump, the Dexcom G6 CGM, and an algorithm built into the pump that adjusts basal insulin delivery and gives automatic correction boluses. The outcomes from this trial were very impressive: Control-IQ helped people spend 2.6 more hours per day in-range (70-180 mg/dl) and lowered A1C by 0.3% more than the group with the same pump and CGM without automation – this is another example where the A1C change is positive, but boy does the time-in-range change explain how much better people did, beyond just the A1C change! These positive results will be submitted to the FDA soon with a US launch (14+ years) expected in 2019, depending on FDA approval. It’s very well agreed that diabetes should be simpler and easier and here’s great news on that front - for those already with the Dexcom G6 and the t:slim X2 pump, getting access to Control-IQ when it is available will be a simple remote software upgrade at home!

Want to know more about the trial? The trial had 168 participants ages 14-71 and followed them for six months. Comparing the Control-IQ hybrid closed loop vs. the same pump and CGM with no automatic correction boluses:

  • Time-in-range (TIR) was 2.6 hours per day better with Control-IQ (70% vs. 59%), with most of the benefit from users spending 2.4 fewer hours per day above 180 mg/dl

  • A1C was 0.3% lower with Control-IQ (7.1% vs. 7.4%)

  • Average blood glucose was 14 mg/dl lower with Control-IQ (156 mg/dl vs. 170 mg/dl)

  • Time below 70 mg/dl was low (less than 2%) in both groups, though Control-IQ did slightly better at reducing time below 70 mg/dl by 13 minutes. That may not sound like so much to everyone but it’s six and a half hours a month less time that participants were hypoglycemic!

With Control-IQ, the time-in-range benefits existed for every level of starting A1C – a positive sign that Control-IQ can help those already at a low A1C and those with an A1C above 8%. Remarkably, all 168 participants completed the study, and Control-IQ users spent 92% of the six months with closed loop active.

In the study questionnaire, people scored Control-IQ:

  • 4.8/5 on desire to continue using

  • 4.7/5 on ease of use

  • 4.6/5 on usefulness

  • 4.5/5 on trust

Control-IQ has no fingerstick CGM calibration (with the Dexcom G6), and no hassle of entering or exiting Auto Mode. These results will be submitted to FDA for Control-IQ to be approved for those aged 14+, though an additional trial for children is also currently recruiting to extend the age range down to six years. 

How does Control-IQ work to keep people in-range?

Control-IQ uses the TypeZero/Dexcom hybrid closed loop algorithm, which still requires all the normal pump settings (basal profile, insulin-to-carb ratio, insulin sensitivity factor. The system aims to keep users between 70-180 mg/dl as much as possible, using a combination of strategies. When the predicted CGM value is:

  • Above 160 mg/dl, it increases the basal insulin rate

  • Above 180 mg/dl, it gives a bolus

  • Below 112.5 mg/dl, it decreases basal insulin

  • Below 70 mg/dl, it turns off basal insulin

Users are still expected to bolus for meals. The automated correction boluses can occur up to once per hour during the day; if the person on the system doesn’t bolus, Control-IQ will “automate” and give approximately 60% of a full correction bolus with a target of 110 mg/dl. These boluses occur on top of the automatic basal rate.

During the day, the system targets a range of 112.5-160 mg/dl. It is more aggressive at night – the goal is to arrive at 112.5-120 mg/dl by the time the user wakes up. Control-IQ also has an “exercise mode” that changes the target to be closer to 140-160 mg/dl.

Dr. Aaron Kowalski, CEO of JDRF and a champion of automated insulin delivery development, stood up at ADA to say, “People will benefit – you see the overnight data, the reduction in hypoglycemia and hyperglycemia. I want to congratulate Tandem and Dexcom, and the competition in this space will provide more options that drive better outcomes.”

For more background on Control-IQ, click here.

What do you think?

About the authors

Adam Brown joined diaTribe in 2010 as a Summer Associate, became Managing Editor in 2011, and served as Senior Editor through 2019. Adam brings almost two decades of personal experience... Read the full bio »